ATEQ ADSE 740 Pitot Static Indicator

 

The ATEQ ADSE 740 Pitot is a complete high performance dual pressure Ps and Pt standalone test bench specially designed to be used in the workshop or in the laboratory to test and calibrate
all air data equipment such as altimeters, vertical speed indicators, air speed indicators, MACH
-meter, air data computers ...) and sensors.


The high precision embedded sensors enable the ATEQ ADSE 740 to be used as a pressure standard.


The man machine interface is programmed under Windows® and Labview®, with a data base managed in a spreadsheet for easy evaluation, management, statistics and presentation.

 
Main Features:
♦ Complete self check of set before use
♦ Selectable pressure units : hPa; mb; in Hg; mmHg; ft; m; kts; km/h, ft/min; m/sec and Mach number
♦ High accuracy, high resolution
♦ Programmable flight envelope to protect equipment under test
♦ RVSM compliant
♦ All four primary flight parameters displayed simultaneously
♦ Programmable leak test
♦Programmable (password write protected) test schedules – 24 programs available
   
Features comparison list:

ATEQ ADSE Pitot Static

tester range

ADSE

712

ADSE

730

ADSE

735

ADSE

740

ADSE

743

ADSE

745

Altimeter testing
Air Speed Indicators Testings
Vertical Speed Indicators Testing

Air Data Computers Testing

Pressure Sensors Testing
Leak Testing

 
Specifications:
General details Measurement specifications
Temperature range operating : 15 °C to 40 °C Function
Range
Accuracy (1)
Power supply 110/240V,50 Hz AC, 50VA Altitude: -2,300 to 80,000ft ±3ft at 0ft
±8ft at 30,000ft
±32ft at 60,000ft

Case:

 

19" x 4 U x 524 mm, 14kg (31lbs)

-700 to 18,000m ±1m at 0m
±2.5m at 9,000m
±10m at 18,000m
Altitude rate: Up to ±15,000ft/min ±1%
max.±5,000m/min ±1%
Screen: 17" LCD colour 2,5kg (5,5 lbs) Indicated airspeed: 10 to 800kts ±2kt at 50kts
±0.14kt at 500kts
±0.07kt at 800kts
Calibration: Recommended period 12 months 20 to 1480km/h ±3km/h at 100km/h
±0.26km/h at 900km/h
±0.13km/h at 1480km/h
Ease of Use

Windows human/machine interface

Program script

Easy programming of test reports

Mach No: 0.1 to 4.0 Mach ±0.002M at 0,8M/25,000ft
±0.004M at 1,7M/30,000ft
Ease of maintenance: Modular design permitting ease of accessibility to mechanical assemblies and electronic components Static sensor 30 to 1200 mbar 0,01% FS
Optional Pitot sensor 30 to 3000 mbar 0,01% FS

Internal pumps

IEEE488 digital interface

Ps & Pt outlet at the back of the bench

Vertical housing

Specific Pt sensor for improved accuracy at low speed

Pneumatic connectors JIC 37 (AN4) or Staubli

     
 

(1) linearity + repeatability + hysteresis at ambiant +10° to +40°C

x 1,5 for -10° to +50°C

x 0,5 for ±2°C lab use